NeuroTrac MyoPlus 2
Dual channel EMG (Biofeedback)
Features
- Dual channel EMG (Biofeedback) combined with NMS-STIM (Neuromuscular Stimulation) & ETS (EMG Triggered Stimulation).
- Passive and Active phases: EMG and NMS in one programme
- Manual and Automatic threshold control.
- EMG Biofeedback between stimulation (muscle activity and tiredness indication when using muscle stimulation).
- Stand alone device or used in conjunction with established multilingual software.
- Pre-set neuromuscular Stimulation modes for: Continence, Sports, Rehab.
- Custom programmes enabling up to 5 phases (any of: STIM W/R-MODCONT- ALT, EMG, ETS) to be set up in each programme.
- Multilingual LCD screen and voice prompts for Biofeedback.
- For home users: Produces and prints (with the optional PC Software) the comprehensive patient progress reports on a day to day basis providing average readings (2 channels) on work/rest, peak, onset and muscle release times, work and rest deviations and the average mA current.
- Used as a learning tool for both patient and clinician with emphasis on improving treatment techniques in the use of EMG and Neuromuscular stimulation.
- Multilingual LCD screen prompts.
- 4 x AA batteries {No mains required}.
The MyoPlus2 is shipped in a hard care complete with all accessories and is ready to use.
PROGRAMMES
PROTOCOLS
- F.E.S. in Stroke
- Treatment of the Hemiplegic Shoulder (Stroke) through Biofeedback: Case Study
- Urinary Incontinence
- Stress Urinary Incontinence, using EMG Triggered Stimulation - NEW
- Pelvic Floor Muscle Activity in Sitting and Standing Postures
- Fecal Incontinence
- Incontinence Case Reports
- Patellofemoral Pain
- 1.1 EMG Range: 0.2 to 2000 ìV RMS (continuous)
- 1.2 Sensitivity: 0.l ìV RMS
- 1.3 Accuracy: 4% of ìV reading +/-0.3 ìV at 200 Hz
- 1.4 Selectable Bandpass filter - 3db Bandwidth, a. Wide: 18 Hz +/- 4 Hz to 370 Hz +/- 10% - Reading below 235 microvolts 10 Hz +/-3 Hz to 370 Hz +/- 10% - Reading above 235 microvolts b. Narrow: 100 Hz +/- 5% to 370 Hz +/- 10%
- 1.5 Notch filter: 50 Hz (Canada 60Hz) - 33 dbs (0. 1% accuracy)
- 1.6 Common Mode Rejection Ratio: 130 dbs Minimum @ 50 Hz
- 1.7 a. Battery: 1.5V, AA battery b. Optional Charger: CE approved.
- 1.8 Work / Rest periods: 2-99 seconds
- 1.9 Number of Trials: 1-99
- 2.1 2 channel Stimulator
- 2.2 Amplitude: 0-90 mA into 500 Ohm load - actual mA will tend to be less than indicated due to Electrode impedance: at 1000 Ohms load (Electrodes in poor condition) the maximum will be limited to 75 mA, at 1500 Ohms load the maximum will be limited to 50 mA.
- 2.3 Type: Constant current, maximum output voltage 70 Volts +5 / -10 Volts
- 2.4 Waveform: Symmetrical, rectangular, bi-phasic with net zero DC current
- 2.5 Pulse width selection: 50-450 ìS (2% accuracy)
- 2.6 Pulse rate selection: 2-100 Hz (2% accuracy)
- 2.7 Work / Rest periods: 2-99 seconds
- 2.8 Time 1 - 99 minutes
- 2.9 Ramp up time: 0.1 - 9.9 seconds
- 2.10 Preset and user programmable treatment Programs
- 2.11 Automatic output shut off with detection of open electrode above 0.5 mA
Battery: Low battery indication at 5.4 volts +/- 0.2 volts, automatic shut off when voltage dropps below the low indication. Replace the batteries immediately! The device goes off automatically when not in use (energy saving): for example when in settings and no key pressed over 1 minute, when in stimulation mode (home screen) and all channels are 0mA.
Environmental Conditions for use: +10 to +30 degrees Centigrade. 0-90% Humidity. Environmental conditions for storage & transport: -10 to +50 degrees Centigrade. 0-90% Humidity. Dimensions: Length 150 mm, Width 89 mm, Depth 35 mm. Weight: MyoPlus2 device: 0.16 Kg (without batteries).