Peritron - Clinical Perineometer

Precision Perineometer for accurate assessment and super sensitive feedback.
Peritron is a hand-held clinical Perineometer intended for assessing the strength of pelvic floor (PF) muscles and teaching pelvic floor exercises. Both vaginal and anal sensors are available.
In operation, air pressure in the sensor caused by a pelvic floor contraction is transferred by a tube to the Readout Unit where it is displayed in several ways.
- Smaller, lighter
- Measuring to 0.1cm/H2O
- More sensitive
- Gives audio as well as visual biofeedback.
- Has linear and logarithmic biofeedback.
- More stats - Gradient, Area under Curve, Peak, Average and Duration
The new Peritron also has a Bluetooth Option* for -
- Saving data for later upload to PC or PDA. This raw data can then be analysed in a spreadsheet program.
Product Information
Use the following links to read detailed information on the new Peritron:
Assessment
Peritron displays the value of the contraction numerically in cm water (cm/H2O). (other units of measurement are available).The value is updated once per 0.6 second during the contraction.Peritron can recall the peak and average values of the contraction and also its duration. It also displays the gradient. Please refer to Quality of the Contraction below.
Normal values
Generally speaking, a vaginally measured contraction value of less than 20cm water indicates that strengthening pelvic floor exercises are advised.Normal anal resting pressures are around 50cm water. Cough usually adds 40 and squeeze 100. These correlate well with traditional measuring techniques using multi-lumen catheters.
^ top
Pressure Profiles
If successive readings are jotted down on a photocopy of the template provided in the Handbook a pressure profile curve may be sketched quickly and easily.
Many Practitioners find this picture of the contraction particularly useful for patient education.
| Rdg | cm water | Rdg | cm water | Rdg | cm water |
| 1 | 20 | 4 | 22 | 7 | 10 |
| 2 | 32 | 5 | 25 | 8 | 1 |
| 3 | 29 | 6 | 17 | 9 |
Readings taken during a contraction
Quality of the Contraction
A strong, sustained contraction is better than a brief, weak one. To help the clinician assess the quality of the contraction, at the press of a button Peritron will recall the peak and average readings as well as the duration of the contraction in seconds. It also displays the gradient.
A STRONG AND SUSTAINED CONTRACTION has similar PEAK and AVERAGE readings. Example vaginal contraction:
AN UNSATISFACTORY CONTRACTION typically has widely differing PEAK and AVERAGE readings. For example:
^ top
Feedback
For teaching pelvic floor contractions the display may be switched to 'progressing bars' that provide 'better-worse' feedback. The more bars that are lit, the stronger the contraction.
|
I I |
I I I I |
I I I I I I |
|
|
|
|
Peritron's super-sensitive response picks up the first 'flickers' of a contraction. This feedback is really appreciated as patients generally need all the encouragement they can get.
Peritron's feedback display has 99 levels of that can be set, that is, degrees of difficulty.
The first, most sensitive range is for learning the exercises; the highest range will challenge the most accomplished pelvic floor athlete! But you can choose anything in between...
^ top
Sensors
The design of our Vaginal and Anal Sensors reflects our many years of experience designing and manufacturing medical products.
- Our sensors may be used with any Perineometer.
- We always welcome inquiries to supply Sensors to other Perineometer Manufacturers on an OEM basis.
- Peritron and PFX use the same Sensors. Thus the sensor of a patient's PFX home exerciser may be brought to the clinic for assessment with Peritron.
- Hygenic and Biocompatible. The contact surfaces of our sensors are of seamless medical grade silicone elastomer, a material chosen for its high biocompatibility. The thin wall of the silicone sheath transmits the pressure faithfully and its performance is temperature independent.
- There are no joins or crevices to trap contaminents. Toxicological testing has shown the sensors to be biologically inert.In contrast, sensors of most other manufacturers have a latex sheath. The FDA require latex products to carry a warning against possible allergic reactions.
- Our Sensors have been approved for safety and efficacy by the USA FDA and carry Class II CE mark classification in Europe.
- Sensitive, repeatable response. The vaginal sensor has a long pressure-responsive area. Consequently, when the sensor is inserted to the recommended depth its pressure sensitive area will span the pelvic floor muscles of 95% of women.This means that, irrespective of who carried out the test, the sensor will be located correctly every time.
- Sensors of most other manufacturers have short pressure-sensitive areas. As a consequence they must be positioned carefully right at the pelvic floor. Variations in positioning leads to session-to-session and operator-to-operator variability in readings.
^ top
Clinical Studies
Those using Peritron in Clinical Research projects need to know its accuracy and repeatability.
The new Peritron is accurate to better than 1 cm H20 throughout the entire range.
With the previous model Peritron, for example, if the reading was 100 then there was a 95% probability that the true pressure is 100 and a 5% probability that it is 99 or 101.
The new Peritron measures to one decimal place (i.e. the scale is different) If it was compared in accuracy (on the same scale) with the old Peritron, it would be 100% accurate. In other words, it is accurate to better than 1 cm H20 throughout the entire range.
To our knowledge Peritron is the most accurate Perineometer currently available.
In practice, the force required to deform the sensor causes a slightly reduced reading. Providing you 'zero' the reading after insertion, the Peritron reading will be about two digits lower than actual applied pressure. (This characteristic affects every Perineometer in the world.) As the sensors are precision mouldings they are all exactly the same and so every reading you take will be about two digits short of the actual pressure indicated. Most regard this discrepancy to be meaningless in a clinical situation.
Collar
In clinical trials it is important to eliminate operator-induced artifacts as far as possible. For this reason we recommend the use of collar cat 2011 that provides repeatable depth of insertion with 100% certainty.
We are pleased to provide the results of our laboratory testing on request.
Inflation Optional for Vaginal (not Anal) Sensors
Most people use the vaginal sensors without inflation. However, a port is provided in the Peritron connecting tube that allows the sensor to be air-inflated by means of a syringe to provide a larger, firmer surface.
A standard inflation pressure of, say, 100 cm provides a constant base-line for readings. The technique used is straight forward:
- Insert sensor
- Inflate to reading of 100
- Press button to zero Peritron reading
- Have patient contract .....
NB: The anal sensor is not intended for inflation.
Single or Multi-Patient Sensors
For such a personal item, most people prefer their own sensor. The modest cost of these makes one-sensor-per-patient a viable clinical policy.
Single-user sensors require only a soap-and-water wash between uses.
^ top
Technical Specifications
Peritron 9300 is a microprocessor-controlled Perineometer that measures the strength of pelvic floor muscle contractions by sensing the pressure of air in a vaginal or anal sensor. The pressure is displayed either numerically in centimetres water pressure or as a multi-range analogue bar-graph.
Peritron is supplied in a moulded case with sensor, connecting tube and handbook. There are two versions:
- Pertiron 9300V incorporates a vaginal sensor
- Peritron 9300A incorporates an anal sensor
- Peritron 9300 is available with both sensors - please refer to ordering page
Operating Range
| Operating Range | 0-300 cm water pressure |
| Accuracy |
|
| Units | Default Units: cmH20
Optional Units
|
| Numerical Readout |
|
| LCD Display |
|
| Analogue Readout | Three-range bar-graph display for exercise feedback |
| Vaginal Sensor of 9300V |
|
| Anal sensor of 9300A |
|
| Carry case | Moulded 23 x 17.5 x 5 cm |
| Auto-off | Function operates after four minutes of no change of pressure |
| Maintenance |
|
| Connectors |
|
| Warranty | Peritron is warranted to be free of defects for 12 months |