Metron Advanced Laser
Specifications
| Mains supply requirements | |
| Voltage | 110 to 240 volts AC |
| Frequecy | 50 - 60 Hz |
| Power | Less than 150 VA |
| Fuses | |
| IEC Mains | 2 of 1A (5 x 20mm) DA205 Delay (T) |
| Secondary internal | 2 of 2A (5 x 20mm) M205 |
| Internal switch mode power supply | |
| Switch mode Power supply unit compliant with EN60601-1:1998 and subsequent amendments. | |
| Secondary voltage | 24 Volts @ 2.7A |
| Laser performance | |
| Output power | Dependant on probe (automatically sensed) Laser Diode Probe 100 mW @ 830nm Optional: SLD/LED Cluster Probe 500mW @ 660/950nm |
| Delivered energy | 0.01 to 99.99 Joules |
| Operation modes | Continuous & pulsed |
| Pulse mode: | |
| Pulse width | 100 uS nominal |
| Pulse frequency | A) Continuous B) 10 Hz, 25 Hz, 50 Hz, 100 Hz, 250 Hz, 500 Hz, 1 kHz, 2 kHz, 5 kHz (Pulses per second) C) Sweep from 10 to 5 kHz (inc continuous) in 10 seconds (1 second each step) |
| Timer | 0 - 99 minutes 59 seconds, 1 sec increments (decrementing). Audible signal and output termination at time expiration |
| Probe (100 mW) | |
| Lasing device | Single GaAs Diode |
| Wavelength | Nominall 810 nm (+/- 10 nm) |
| Power | 100 mW (+/- 10 mW) |
| Treatment area illumination | 3 x Blue LED's (470 nm, visible through eyewear protection that attenuates Infrared/Near infrared) |
| Output Activation | Capacitance switch on laser probe handle |