EMG-EZ Biofeedback Machine Model 4120
Single Channel surface EMG System
Main Features & Benefits
- Two modes of biofeedback - LED and 5-way audible tone
- 5-way audible tone feedback includes Proportional, Above and Below Thresholds and 3-second delayed alarm
- 13 segment LED bargraph visual feedback
- Wide and narrow filters provide precise monitoring of normal and fatiguing muscles and eliminate EKG artifact
- Highly-sensitive active sensor needs no skin preparation to locate muscle sites quickly
- DC voltage output connects accessories and permits computer monitoring
Included in the package
- Two leads
- Nylatex® wrap
- Wrist strap
- Belt clip
- Carrying case
- Four disposable electrodes
- Earphones
- Batteries
- Instruction manual.
Product 'Quick Start' Guide
illustrated tech-note will aid in the fi rst time use of the EMG-EZ single channel sEMG Biofeedback System. It is not a substitute for the full user manual, which should be read and thoroughly understood.
EMG Electrodes
We supply the Dura-Stick Snap EMG electrodes in either singles or a packet of 100
5cm Round, White. Also suitable for Myotrac EMG
Also available
- 41017 ·Sensor
- 41015 ·Sensor Extender Cable
- 41016 ·Earphone
- 77702 ·Pin to Snap Extension Adapters
- 77706 ·Snap to Snap Extension Adapters
EMG Specifications
| Input Impedance: | 1,000,000 MΩ in parallel with 10 pF |
| Signal Input Range x1: | 0-20 µV |
| Signal Input Range x10: | 0-200 µV |
| Signal Input Range x100: | 0-2000 µV |
| CMRR 20 - 500 Hz (approx.): | 130 dB |
| CMRR @ 60 Hz (approx.): | 180 dB |
| Channel Bandwidth (Wide): | 20 Hz ±5 Hz-500 Hz ±50 Hz |
| Channel Bandwidth (Narrow): | 100 Hz ±10 Hz-200 Hz ±20 Hz |
| Signal Output Range: | 0 V-2 V |
| Supply Voltage: | 7.4 V-9.6 V |
| Current Consumption: | 10 mA-3 mA /+5 mA |
| Battery Life (Alkaline): | 40 Hours minimum |
| Low Battery Warning: | 7.4 V |
| Accuracy: | ±5% and ±0.3 mV |
| Electrical Safety Class: | Class 1, Type BF |
| Dimensions: | (L x W x H) 1" x 2.4" x 4.4" (25 mm x 61 mm x 112 mm) |
| Weight: | 2.5 oz (71 g) |
| Safety Tests: | UL/IEC/EN 60601-1,IEC/EN 60601-1-2, IEC 60601-2-10 |
 